Dimensions: 420 mm wide x 270 mm deep x 210 mm high
Weight: 15.5 kg
Operating environment: For indoor use only
Operating temperature: +5 °C to +40 °C for safe operation. See also temperature control range restriction.
Operating humidity: 5% to 90% relative humidity non-condensing
Altitude: up to 2000 m
Pollution degree 2 (BS EN61010-1)
Temperature control range: (ambient + 5 °C) to 40 °C.
Temperature measurement accuracy: ± 0.2 °C
Temperature control accuracy: ± 0.1 °C measured after any transient effects due to set-point changes have subsided.
Flow control range: 0 ml/minute to 900 ml/minute.
Flow measurements are normalised to 0 C , 50% RH and 1 bar.
Flow accuracy: The greater of ± 10% or ± 0.3 ml/minute
Flow control accuracy: The greater of ± 5% or ± 0.2 ml/minute measured after any transient effects due to set-point changes have subsided.
Capacity / Dishes per chamber:
4 x NUNC 4 well dishes
4 x NUNC 60 mm dishes
10 x NUNC 30 mm dishes
4 x MINITUB 5 well dishes
4 x FALCON 60 mm dishes
4 x FALCON 60mm single - well “organ culture” dishes
Power requirements includes Controller 100 - 230 V~ / 50/60Hz / 1.1 A
Internal battery backup: Gelled sealed lead acid battery / 12 v x 12 Ah
Gas supply: Premixed gas. Typically 6% CO2, 5% O2, balance N2
Supply pressure: 1.5 ± 0.15 bar
Connectors: SWAGELOK 1/4” tube fitting
The incubator provides 3 volt-free terminals which provide normally-open and normally-closed contacts.
Remote monitoring via LAN
10 Base T Ethernet - RJ45 shielded. Modbus-TCP-IP protocol.
Remote PT100 sensors: PT100 Class A to EN60751.
In order to fit the monitoring ports the sensor must meet the following specification:
Maximum diameter 2.51 mm.
Minimum length 100 mm.
Sensing region should be within 15 mm of the tip.